Pumps

EPSIMED ZNB-XAII Infusion Pump
Roll over image to zoom
Product images:

 

 

Model: Infusion Pump EPSIMED ZNB-XAII

 

 

Pumping Mechanism Curvilinear peristaltic

 

 

IV Set Compatible with IV sets of any standard

 

 

Flow Rate 0.1-1500 ml/h (in 0.1 ml/h increments)

 

 

Purge, Bolus 100-1500 ml/h (in 0.1 m/h increments)

 

 

Purge when pump stops, bolus when pump starts,

 

 

Accuracy ±3%

 

 

VTBI 1-20000 ml (in 0.1 ml increments)

 

 

Infusion Mode ml/h, drop/min, time-based, body weight, nutrition

 

 

KVO Rate 0.1-5 ml/h (in 0.1 ml/h increments)

 

 

Alarms Occlusion, air-in-line, door open, end program, low

 

 

battery, end battery, AC power off, motor malfunction,

 

 

system malfunction, reminder alarm

 

 

Additional Features Real-time infused volume, automatic power switching,

 

 

mute key, purge, bolus, system memory, history log,

 

 

drug library, key locker, rotary knob, change flow rate

 

 

without stopping the pump

 

 

Occlusion Sensitivity High, medium, low

 

 

History Log 50000 events

 

 

Air-in-line Detection Ultrasonic detector

 

 

Wireless management Optional

 

 

Drop Sensor Optional

 

 

Vehicle Power (Ambulance) 12±1.2 V

 

 

Power Supply, AC 110/230 V (optional), 50-60 Hz, 20 VA

 

 

Battery 9.6±1.6 V, rechargeable

 

 

Battery Life 5 hours at 25 ml/h

 

 

Working Temperature 10-30 degrees

 

 

Relative Humidity 30-75%

 

 

Atmospheric Pressure 860-1060 hpa

 

 

Size 130*145*228 mm

 

 

Weight 2.5 kg

 

 

Safety Classification Class

Availability:
In Stock
EPSIMED ZNB-XAII Infusion Pump

 

 

Model: Infusion Pump EPSIMED ZNB-XAII

 

 

Pumping Mechanism Curvilinear peristaltic

 

 

IV Set Compatible with IV sets of any standard

 

 

Flow Rate 0.1-1500 ml/h (in 0.1 ml/h increments)

 

 

Purge, Bolus 100-1500 ml/h (in 0.1 m/h increments)

 

 

Purge when pump stops, bolus when pump starts,

 

 

Accuracy ±3%

 

 

VTBI 1-20000 ml (in 0.1 ml increments)

 

 

Infusion Mode ml/h, drop/min, time-based, body weight, nutrition

 

 

KVO Rate 0.1-5 ml/h (in 0.1 ml/h increments)

 

 

Alarms Occlusion, air-in-line, door open, end program, low

 

 

battery, end battery, AC power off, motor malfunction,

 

 

system malfunction, reminder alarm

 

 

Additional Features Real-time infused volume, automatic power switching,

 

 

mute key, purge, bolus, system memory, history log,

 

 

drug library, key locker, rotary knob, change flow rate

 

 

without stopping the pump

 

 

Occlusion Sensitivity High, medium, low

 

 

History Log 50000 events

 

 

Air-in-line Detection Ultrasonic detector

 

 

Wireless management Optional

 

 

Drop Sensor Optional

 

 

Vehicle Power (Ambulance) 12±1.2 V

 

 

Power Supply, AC 110/230 V (optional), 50-60 Hz, 20 VA

 

 

Battery 9.6±1.6 V, rechargeable

 

 

Battery Life 5 hours at 25 ml/h

 

 

Working Temperature 10-30 degrees

 

 

Relative Humidity 30-75%

 

 

Atmospheric Pressure 860-1060 hpa

 

 

Size 130*145*228 mm

 

 

Weight 2.5 kg

 

 

Safety Classification Class