Pumps

EPSIMED ZNB-XAII Infusion Pump
Roll over image to zoom
Product images:

Model: Infusion Pump EPSIMED ZNB-XAII

Pumping Mechanism Curvilinear peristaltic

IV Set Compatible with IV sets of any standard

Flow Rate 0.1-1500 ml/h (in 0.1 ml/h increments)

Purge, Bolus 100-1500 ml/h (in 0.1 m/h increments)

Purge when pump stops, bolus when pump starts,

Accuracy ±3%

VTBI 1-20000 ml (in 0.1 ml increments)

Infusion Mode ml/h, drop/min, time-based, body weight, nutrition

KVO Rate 0.1-5 ml/h (in 0.1 ml/h increments)

Alarms Occlusion, air-in-line, door open, end program, low

battery, end battery, AC power off, motor malfunction,

system malfunction, reminder alarm

Additional Features Real-time infused volume, automatic power switching,

mute key, purge, bolus, system memory, history log,

drug library, key locker, rotary knob, change flow rate

without stopping the pump

Occlusion Sensitivity High, medium, low

History Log 50000 events

Air-in-line Detection Ultrasonic detector

Wireless management Optional

Drop Sensor Optional

Vehicle Power (Ambulance) 12±1.2 V

Power Supply, AC 110/230 V (optional), 50-60 Hz, 20 VA

Battery 9.6±1.6 V, rechargeable

Battery Life 5 hours at 25 ml/h

Working Temperature 10-30 degrees

Relative Humidity 30-75%

Atmospheric Pressure 860-1060 hpa

Size 130*145*228 mm

Weight 2.5 kg

Safety Classification Class

Availability:
In Stock
EPSIMED ZNB-XAII Infusion Pump

Model: Infusion Pump EPSIMED ZNB-XAII

Pumping Mechanism Curvilinear peristaltic

IV Set Compatible with IV sets of any standard

Flow Rate 0.1-1500 ml/h (in 0.1 ml/h increments)

Purge, Bolus 100-1500 ml/h (in 0.1 m/h increments)

Purge when pump stops, bolus when pump starts,

Accuracy ±3%

VTBI 1-20000 ml (in 0.1 ml increments)

Infusion Mode ml/h, drop/min, time-based, body weight, nutrition

KVO Rate 0.1-5 ml/h (in 0.1 ml/h increments)

Alarms Occlusion, air-in-line, door open, end program, low

battery, end battery, AC power off, motor malfunction,

system malfunction, reminder alarm

Additional Features Real-time infused volume, automatic power switching,

mute key, purge, bolus, system memory, history log,

drug library, key locker, rotary knob, change flow rate

without stopping the pump

Occlusion Sensitivity High, medium, low

History Log 50000 events

Air-in-line Detection Ultrasonic detector

Wireless management Optional

Drop Sensor Optional

Vehicle Power (Ambulance) 12±1.2 V

Power Supply, AC 110/230 V (optional), 50-60 Hz, 20 VA

Battery 9.6±1.6 V, rechargeable

Battery Life 5 hours at 25 ml/h

Working Temperature 10-30 degrees

Relative Humidity 30-75%

Atmospheric Pressure 860-1060 hpa

Size 130*145*228 mm

Weight 2.5 kg

Safety Classification Class